  • Abstract
    Comparisons are made between the average magnetic flux density as would be measured with a three-axis coil probe and the flux density at the center of the probe. Probability distributions of the differences between the two quantities are calculated assuming a dipole magnetic field and are found to be asymmetric. The distributions allow estimates of uncertainty for resultant magnetic field measurements made near some electrical appliances and other electrical equipment
  • Keywords
    error analysis; magnetic field measurement; magnetometers; measurement errors; probability; asymmetric distributions; average magnetic flux density; dipole magnetic field; electrical appliances; electrical equipment; estimates of uncertainty; flux density; magnetic field measurements; measurement error; probability distributions; three-axis coil probe; three-axis magnetic field meters; Coils; Displays; Electrical products; Magnetic field measurement; Magnetic fields; Magnetic flux; Magnetic flux density; Measurement errors; Probes; Statistical distributions;