DocumentCode
1234857
Title
Distributions of measurement error for three-axis magnetic field meters during measurements near appliances
Author
Misakian, Martin ; Fenimore, Charles
Author_Institution
Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
Volume
45
Issue
1
fYear
1996
fDate
2/1/1996 12:00:00 AM
Firstpage
244
Lastpage
249
Abstract
Comparisons are made between the average magnetic flux density as would be measured with a three-axis coil probe and the flux density at the center of the probe. Probability distributions of the differences between the two quantities are calculated assuming a dipole magnetic field and are found to be asymmetric. The distributions allow estimates of uncertainty for resultant magnetic field measurements made near some electrical appliances and other electrical equipment
Keywords
error analysis; magnetic field measurement; magnetometers; measurement errors; probability; asymmetric distributions; average magnetic flux density; dipole magnetic field; electrical appliances; electrical equipment; estimates of uncertainty; flux density; magnetic field measurements; measurement error; probability distributions; three-axis coil probe; three-axis magnetic field meters; Coils; Displays; Electrical products; Magnetic field measurement; Magnetic fields; Magnetic flux; Magnetic flux density; Measurement errors; Probes; Statistical distributions;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.481341
Filename
481341
Link To Document