The Use of a Damping Chamber and Sine Wave Oscillator for Optimal Frequency Response in Pressure Recording
Author :
Jennings, Rufus B., Jr. ; Krovetz, L.Jerome
Author_Institution :
Department of Pediatrics, The Johns Hopkins University School of Medicine and the Children´´s Medical and Surgical Center, The Johns Hopkins Hospital, Baltimore, Maryland 21205.
Issue :
2
fYear :
1970
fDate :
4/1/1970 12:00:00 AM
Firstpage :
134
Lastpage :
137
Abstract :
Pressure recording is frequently performed with catheter-manometer systems having high frequency response but inadequate damping. Resonance amplification of catheter movement artefact is often a problem with underdamped systems. The amount of amplitude amplification at the resonant frequency of a water-filled catheter and transducer is a function of the damping of the system.
Keywords :
Catheters; Damping; Fasteners; Frequency response; Impedance; Oscillators; Pediatrics; Resonance; Resonant frequency; Transducers;