A new X-ray microcalorimeter based on a pixelated TES array
Author :
Fukuda, Daiji ; Ohno, Masashi ; Takahashi, Hiroyuki ; Kunieda, Yuichi ; Inou, Tadashi ; Ohkubo, Masataka ; Ataka, Manabu ; Nakazawa, Masaharu
Author_Institution :
Univ. of Tokyo, Japan
Volume :
13
Issue :
2
fYear :
2003
fDate :
6/1/2003 12:00:00 AM
Firstpage :
653
Lastpage :
656
Abstract :
We are developing a new x-ray microcalorimeter based on superconducting transition edge sensors (TES) as an imaging sensor. This device has ten pixelated transition edge sensors with Iridium superconductive films. When a constant bias voltage is applied to all pixels, each pixel is operated at slightly different equilibrium temperature. This arises from the different thermal responses between pixels, so that response signal shapes would vary according to the position of the incident x-ray. We have fabricated a prototype of the pixelated array and examined its performance. The position dependency measurements by scanning the collimated x-ray over the device have successfully shown that the device is able to resolve its pixel position. The energy resolution of a test device was 13.1 eV (FWHM) for 3 keV x-rays.
Keywords :
X-ray detection; image sensors; iridium; particle calorimetry; position sensitive particle detectors; superconducting arrays; superconducting particle detectors; superconducting thin films; 3 keV; Ir; X-ray microcalorimeter; imaging sensor; iridium superconductive film; pixelated TES array; position sensitive detector; superconducting transition edge sensor; Energy resolution; Image sensors; Optical imaging; Prototypes; Shape; Superconducting films; Superconductivity; Temperature sensors; Voltage; X-ray imaging;
fLanguage :
English
Journal_Title :
Applied Superconductivity, IEEE Transactions on
Publisher :
ieee
ISSN :
1051-8223
Type :
jour
DOI :
10.1109/TASC.2003.813986
Filename :
1211687
Link To Document :
بازگشت