DocumentCode :
1242000
Title :
A 256 x 256 2-D array transducer with row-column addressing for 3-D rectilinear imaging
Author :
Seo, Chi Hyung ; Yen, Jesse T.
Author_Institution :
Dept. of Bioeng., Univ. of Washington, Seattle, WA
Volume :
56
Issue :
4
fYear :
2009
fDate :
4/1/2009 12:00:00 AM
Firstpage :
837
Lastpage :
847
Abstract :
We present simulation and experimental results from a 5-MHz, 256times256 2-D (65536 elements, 38.4times38.4 mm) 2-D array transducer with row-column addressing. The main benefits of this design are a reduced number of interconnects, a modified transmit/receive switching scheme with a simple diode circuit, and an ability to perform volumetric imaging of targets near the transducer with transmit beamforming in azimuth and receive beamforming in elevation. The final dimensions of the transducer were 38.4 mm times 38.4 mm times 300 mum. After a row-column transducer was prototyped, the series resonance impedance was 104 Omega at 5.4 MHz. The measured -6 dB fractional bandwidth was 53% with a center frequency of 5.3 MHz. The SNR at the transmit focus was measured to be 30 dB. At 5 MHz, the average nearest neighbor crosstalk was -25 dB. In this paper, we present 3-D images of both 5 pairs of nylon wires embedded in a clear gelatin phantom and an 8 mm diameter cylindrical anechoic cyst phantom acquired from a 256 times 256 2-D array transducer made from a 1-3 composite. We display the azimuth and elevation B-scans as well as the C-scan for each image. The cross-section of the wires is visible in the azimuth B-scan, and the long axes can be seen in the elevation B-scan and C-scans. The pair of wires with 1-mm axial separation is discernible in the elevational B-scan. When a single wire from the wire target phantom was used, the measured lateral beamwidth was 0.68 mm and 0.70 mm at 30 mm depth in transmit beamforming and receive beamforming, respectively, compared with the simulated beamwidth of 0.55 mm. The cross-section of the cyst is visible in the azimuth B-scan whereas the long axes can be seen as a rectangle in the elevation B-scan and C-scans.
Keywords :
biomedical transducers; polymers; ultrasonic imaging; ultrasonic transducers; 2-D array transducer; 3-D rectilinear imaging; cylindrical anechoic cyst phantom; depth 30 mm; frequency 5.4 MHz; gelatin phantom; nylon wires; row-column addressing; simple diode circuit; size 8 mm; transmit/receive switching scheme; Array signal processing; Azimuth; Circuit simulation; Diodes; Imaging phantoms; Integrated circuit interconnections; Prototypes; Switching circuits; Transducers; Wires; Computer-Aided Design; Equipment Design; Equipment Failure Analysis; Image Enhancement; Reproducibility of Results; Sensitivity and Specificity; Signal Processing, Computer-Assisted; Transducers; Ultrasonography;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/TUFFC.2009.1107
Filename :
4815314
Link To Document :