• DocumentCode
    1243035
  • Title

    Impairment-Based 3-D Robotic Intervention Improves Upper Extremity Work Area in Chronic Stroke: Targeting Abnormal Joint Torque Coupling With Progressive Shoulder Abduction Loading

  • Author

    Ellis, Michael D. ; Sukal-Moulton, Theresa M. ; Dewald, Julius P A

  • Author_Institution
    Dept. of Phys. Therapy & Human Movement Sci., Northwestern Univ., Chicago, IL
  • Volume
    25
  • Issue
    3
  • fYear
    2009
  • fDate
    6/1/2009 12:00:00 AM
  • Firstpage
    549
  • Lastpage
    555
  • Abstract
    The implementation of a robotic system ( ACT3D ) that allowed for a quantitative measurement of abnormal joint torque coupling in chronic stroke survivors and, most importantly, a quantitative means of initiating and progressing an impairment-based intervention, is described. Individuals with chronic moderate to severe stroke (n = 8) participated in this single-group pretest-posttest design study. Subjects were trained over eight weeks by progressively increasing the level of shoulder abduction loading experienced by the participant during reaching repetitions as performance improved. Reaching work area was evaluated pre- and postintervention for ten different shoulder abduction loading levels along with isometric single-joint strength and a qualitative clinical assessment of impairment. There was a significant effect of session (pre versus post) with an increase in reaching work area, despite no change in single-joint strength. This data suggests that specifically targeting the abnormal joint torque coupling impairment through progressive shoulder abduction loading is an effective strategy for improving reaching work area following hemiparetic stroke. Application of robotics, namely, the ACT3D , allowed for quantitative control of the exercise parameters needed to directly target the synergistic coupling impairment. The targeted reduction of abnormal joint torque coupling is likely the key factor explaining the improvements in reaching range of motion achieved with this intervention.
  • Keywords
    biomechanics; handicapped aids; medical robotics; patient rehabilitation; abnormal joint torque coupling; chronic stroke; impairment-based 3D robotic intervention; movement impairment; progressive shoulder abduction loading; rehabilitation robotic system; upper extremity work area; Arm; kinematics; kinetics; rehabilitation robotics; stroke;
  • fLanguage
    English
  • Journal_Title
    Robotics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1552-3098
  • Type

    jour

  • DOI
    10.1109/TRO.2009.2017111
  • Filename
    4815505