• DocumentCode
    1244369
  • Title

    Coherent plane-wave compounding for very high frame rate ultrasonography and transient elastography

  • Author

    Montaldo, Gabriel ; Tanter, Mickaël ; Bercoff, Jérémy ; Benech, Nicolas ; Fink, Mathias

  • Author_Institution
    Ecole Super. de Phys. et Chimie Industrielles (ESPCI), INSERM, Paris
  • Volume
    56
  • Issue
    3
  • fYear
    2009
  • fDate
    3/1/2009 12:00:00 AM
  • Firstpage
    489
  • Lastpage
    506
  • Abstract
    The emergence of ultrafast frame rates in ultrasonic imaging has been recently made possible by the development of new imaging modalities such as transient elastography. Data acquisition rates reaching more than thousands of images per second enable the real-time visualization of shear mechanical waves propagating in biological tissues, which convey information about local viscoelastic properties of tissues. The first proposed approach for reaching such ultrafast frame rates consists of transmitting plane waves into the medium. However, because the beamforming process is then restricted to the receive mode, the echographic images obtained in the ultrafast mode suffer from a low quality in terms of resolution and contrast and affect the robustness of the transient elastography mode. It is here proposed to improve the beamforming process by using a coherent recombination of compounded plane-wave transmissions to recover high-quality echographic images without degrading the high frame rate capabilities. A theoretical model is derived for the comparison between the proposed method and the conventional B-mode imaging in terms of contrast, signal-to-noise ratio, and resolution. Our model predicts that a significantly smaller number of insonifications, 10 times lower, is sufficient to reach an image quality comparable to conventional B-mode. Theoretical predictions are confirmed by in vitro experiments performed in tissue-mimicking phantoms. Such results raise the appeal of coherent compounds for use with standard imaging modes such as B-mode or color flow. Moreover, in the context of transient elastography, ultrafast frame rates can be preserved while increasing the image quality compared with flat insonifications. Improvements on the transient elastography mode are presented and discussed.
  • Keywords
    biological tissues; biomedical ultrasonics; phantoms; beamforming process; coherent plane-wave compounding; coherent recombination; compounded plane-wave transmissions; conventional B-mode imaging; echographic images; insonifications; tissue-mimicking phantoms; transient elastography; ultrafast frame; ultrasonography; Array signal processing; Biological tissues; Data acquisition; Data visualization; Elasticity; Image quality; Image resolution; Ultrasonic imaging; Ultrasonography; Viscosity; Algorithms; Elasticity; Elasticity Imaging Techniques; Female; Humans; Image Enhancement; Models, Statistical; Models, Theoretical; Signal Processing, Computer-Assisted; Ultrasonography; Ultrasonography, Mammary; Viscosity;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/TUFFC.2009.1067
  • Filename
    4816058