• DocumentCode
    1251795
  • Title

    Primary calibration of hydrophones with extended frequency range 1 to 70 MHz using optical interferometry

  • Author

    Koch, Christian ; Molkenstruck, Walter

  • Author_Institution
    Phys. Tech. Bundesanstalt, Braunschweig, Germany
  • Volume
    46
  • Issue
    5
  • fYear
    1999
  • Firstpage
    1303
  • Lastpage
    1314
  • Abstract
    A technique for the primary calibration of hydrophones based on an optical principle is presented. An interferometer determines the displacement of a pellicle mounted on the surface of the tank fluid perpendicular to the sound propagation axis; the known acoustic field is then measured using the hydrophone. The application range extends from 1 to 50 MHz and, with a lesser accuracy, to 70 MHz. The loaded and the open-circuit sensitivity are obtained pointwise, and the frequency and the number of the measurement points can be arbitrarily chosen. As an example, two membrane hydrophones with a spot diameter of 0.5 mm are calibrated, and the comparison with a theoretical model describing the acoustic and electrical properties of the hydrophone shows good agreement.
  • Keywords
    acoustic field; calibration; displacement measurement; hydrophones; light interferometry; measurement by laser beam; 1 to 70 MHz; acoustic field; displacement measurement; extended frequency range; fluid surface mounted pellicle; hydrophone calibration; laser interferometry; membrane hydrophones; model; open-circuit sensitivity; optical interferometry; primary calibration; uncertainties; Acoustic measurements; Acoustic propagation; Biomembranes; Calibration; Displacement measurement; Frequency measurement; Optical interferometry; Optical propagation; Optical sensors; Sonar equipment;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/58.796135
  • Filename
    796135