DocumentCode
1252059
Title
Development, Calibration, and Experimental Results Obtained With an Innovative Calorimeter (CALMOS) for Nuclear Heating Measurements
Author
Carcreff, Hubert ; Clouté-Cazalaa, Véronique ; Salmon, Laurent
Author_Institution
CEA/DEN/ DRSN/SIREN/LASPI (Saclay), Gif sur Yvette Cedex, France
Volume
59
Issue
4
fYear
2012
Firstpage
1369
Lastpage
1376
Abstract
Nuclear heating inside an MTR reactor has to be known in order to be able to control samples temperature during irradiation experiments. An R&D program has been carried out at CEA to design a new type of in-core calorimetric system. This new development, started in 2002, has for main objective to manufacture a calorimeter suitable to monitoring nuclear heating inside the 70 MWth OSIRIS material testing reactor operated by CEA´s Nuclear Energy Division at the Saclay research center. An innovative calorimetric probe, associated to a specific handling system, has been designed to provide access to measurements both along the fissile height and on the upper part of the core, where nuclear heating still remains high. Two mock-ups of the probe were manufactured and tested in 2005 and 2009 in ex-core area of OSIRIS reactor for process validation, while a displacement system has been especially studied to move the probe along a given axial measurement range. This paper deals with the development, tests on preliminary mock-ups and the finalization of the probe. Main modeling and experimental results are presented. Moreover, alternative methods to calibration for nuclear heating rate measurements which are now possible with this new calorimeter are presented and discussed.
Keywords
Aluminum; Calibration; Inductors; Probes; Resistance heating; Temperature measurement; Differential calorimeter; heating rate; nuclear heating;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2012.2207740
Filename
6249778
Link To Document