• DocumentCode
    1252597
  • Title

    Measurement error for six port ANA

  • Author

    Bergeault, E. ; Huyart, B. ; Geneves, G. ; Jallet, L.

  • Author_Institution
    Lab. Central des Ind. Electr., Fontenay Aux Roses, France
  • Volume
    137
  • Issue
    6
  • fYear
    1990
  • fDate
    12/1/1990 12:00:00 AM
  • Firstpage
    415
  • Lastpage
    416
  • Abstract
    Although six-port ANA theory requires that the input impedances of power detectors remain constant with incident power, this is not generally the case if diode detectors are used. Consequently, uncertainties in the measurement of the unknown load reflection coefficient appear. Measurements of the variations of diode impedances are used in a computer simulation to estimate these uncertainties.
  • Keywords
    calibration; computerised instrumentation; electric impedance; measurement errors; microwave reflectometry; network analysers; 2 GHz; automatic network analyser; computer simulation; diode detectors; diode impedances; load reflection coefficient; measurement error; power detectors; six port ANA; six port reflectometer;
  • fLanguage
    English