The design of a hand-held K-factor meter using an air-core current transformer
Author
Kung, Chih-Hsien ; Devaney, Michael J. ; Huang, Chung-Ming
Author_Institution
Dept. of Inf. Manage., Chang-Jung Univ., Tainan, Taiwan
Volume
46
Issue
4
fYear
1997
fDate
8/1/1997 12:00:00 AM
Firstpage
811
Lastpage
816
Abstract
This paper describes in detail the design of a portable microcontroller-based K-factor meter, or transformer harmonic derating factor meter. This innovative instrument is based on the recently developed time domain method for K-factor determination. K-factor is measured from the ratio of the rms value of the time derivative of a current to the rms value of this same current. The meter employs the properties of an air-core current transformer, or Rogowski coil, to acquire the time derivative of a current and then numerical integration, of this digitized signal, to obtain the current waveform itself. This method avoids the noise of the numeric differentiation, characteristic of the time domain approach to K-factor measurement, as well as the computational effort of the fast Fourier transform in the frequency domain approach. In addition, the instrument incorporates thermal demand models for both K-factor and rms current to assess transformer harmonic heating over a user selected time interval. A prototype meter, using a MC68HC16 microcontroller, was built and tested with a variety of actual and synthesized power system waveforms
Keywords
automatic test equipment; coils; current transformers; harmonic analysis; portable instruments; power engineering computing; power transformer testing; power transformers; MC68HC16 microcontroller; Rogowski coil; air-core current transformer; current waveform; digitized signal; eddy current; frequency domain approach; hand-held K-factor meter; harmonic heating; numeric differentiation; numerical integration; power quality fast Fourier transform; rms current; synthesized power system waveforms; thermal demand models; time derivative; time domain method; time interval; transformer harmonic derating factor meter; Coils; Current measurement; Current transformers; Fast Fourier transforms; Frequency measurement; Instruments; Noise measurement; Power system harmonics; Power system modeling; Time measurement;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.650779
Filename
650779
Link To Document