Abstract
This paper particularly introduces error model, theory, algorithm of the unknown thru calibration technique, and shows why the thru standard has an approximate delay expression. Experiment for coaxial non-insertable device has been processed showing that the unknown thru calibration technique has a great increase in the measurement accuracy and consistency.
Keywords
approximation theory; calibration; measurement errors; measurement standards; approximate delay expression; coaxial noninsertable device; measurement error model; unknown thru calibration technique; Calibration; Facsimile; Measurement uncertainty; Metrology; Ports (Computers); Reflection; Standards;