DocumentCode :
125572
Title :
Study on the unknown thru calibration technique
Author :
Zhang Na ; Zhang Guo Hua ; Chen Ting ; Liu Jie
Author_Institution :
Beijing Inst. of Radio Metrol. & Meas., Beijing, China
fYear :
2014
fDate :
16-23 Aug. 2014
Firstpage :
1
Lastpage :
4
Abstract :
This paper particularly introduces error model, theory, algorithm of the unknown thru calibration technique, and shows why the thru standard has an approximate delay expression. Experiment for coaxial non-insertable device has been processed showing that the unknown thru calibration technique has a great increase in the measurement accuracy and consistency.
Keywords :
approximation theory; calibration; measurement errors; measurement standards; approximate delay expression; coaxial noninsertable device; measurement error model; unknown thru calibration technique; Calibration; Facsimile; Measurement uncertainty; Metrology; Ports (Computers); Reflection; Standards;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
General Assembly and Scientific Symposium (URSI GASS), 2014 XXXIth URSI
Conference_Location :
Beijing
Type :
conf
DOI :
10.1109/URSIGASS.2014.6929005
Filename :
6929005
Link To Document :
بازگشت