DocumentCode :
1260066
Title :
High precision measurement of DC voltage ratios from 20 V/10 V to 1000 V/10 V
Author :
Zhang, Xiuzeng ; Qie, Jiaping ; Zhang, Lili ; Shao, Haiming
Author_Institution :
Electromagn. Div., Nat. Inst. of Metrol., Beijing, China
Volume :
51
Issue :
1
fYear :
2002
fDate :
2/1/2002 12:00:00 AM
Firstpage :
59
Lastpage :
62
Abstract :
This paper presents a new method of calibrating de voltage ratios from 20 V: 10 V to 1000 V: 10 V. A new battery-powered differential voltage detector has been developed to reduce the uncertainty of the measurement system. An absolute self-calibration process was used and traceability to a voltage ratio standard is not necessary. The uncertainties of dc voltage ratios (20 V to 1000 V): 10 V were less than 2 × 10-7
Keywords :
calibration; stability; voltage dividers; voltage measurement; 1000 V to 10 V; 1000 to 10 V; 20 V to 10 V; 20 to 10 V; DC voltage divider; DC voltage ratio; absolute self-calibration; battery-powered differential voltage detector; reference divider; traceability; uncertainties; uncertainty; voltage ratio standard; Calibration; Capacitors; Circuits; Detectors; Preamplifiers; Resistors; Sampling methods; Stability; Temperature; Voltage measurement;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/19.989900
Filename :
989900
Link To Document :