Title :
Tests on the IEN electrometric AC-DC transfer standard
Author :
Pogliano, Umberto ; Bosco, Gian Carlo
Author_Institution :
Ist. Elettrotecnico Nazionale Galileo Ferraris, Torino, Italy
fDate :
2/1/2002 12:00:00 AM
Abstract :
Tests on the electrometric AC-DC transfer standard built at IEN have been performed. The instrument was shown to work correctly, and comparison with thermal converters showed a short-term repeatability (1σ) of about 2 parts in 106. Mechanical effects such as deformations and vibrations have been analyzed, both theoretically and experimentally. Direct determinations of the transfer difference made using thermal converters have shown results almost comparable with those obtained by the step-up procedure. The long-term stability of the instrument has also been investigated
Keywords :
deformation; electrometers; stability; transfer standards; vibrations; voltage measurement; IEN electrometric AC-DC transfer standard tests; deformations; electrometric AC-DC transfer standard; long-term instrument stability; mechanical effects; short-term repeatability; step-up procedure; thermal converters; transfer difference; vibrations; Electrodes; Force measurement; Gold; Instruments; Measurement standards; Prototypes; Shafts; Testing; Vibrations; Voltage;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on