DocumentCode :
1267616
Title :
A calibration-free method for measurement of sound speed in biological tissue samples
Author :
Ophir, J. ; Lin, T.
Author_Institution :
Dept. of Radiol., Texas Univ., Med. Sch., Houston, TX, USA
Volume :
35
Issue :
5
fYear :
1988
Firstpage :
573
Lastpage :
577
Abstract :
A transmission method for speed-of-sound measurement in biological tissue specimens is described. The technique involves detecting the time of arrival of a transmitted pulse with a small hydrophone that penetrates the tissue sample in accurately known increments. The technique allows localized measurements to be made without the error-prone task of measuring the thickness of the specimen and without calibration of the apparatus in a reference medium with known speed of sound.<>
Keywords :
bioacoustics; biological techniques and instruments; ultrasonic velocity measurement; biological tissue; calibration-free method; hydrophone; sound speed; transmission method; ultrasonic velocity; Acoustic pulses; Acoustic transducers; Biological materials; Biological tissues; Calibration; Pulse measurements; Sonar equipment; Thickness measurement; Time measurement; Velocity measurement;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/58.8035
Filename :
8035
Link To Document :
بازگشت