DocumentCode :
1278468
Title :
Techniques for development of safety-related software for surgical robots
Author :
Varley, Peter
Author_Institution :
Dept. of Comput. Sci., Cardiff Univ., UK
Volume :
3
Issue :
4
fYear :
1999
Firstpage :
261
Lastpage :
267
Abstract :
Regulatory bodies require evidence that software controlling potentially hazardous devices is developed to good manufacturing practices. Effective techniques used in other industries assume long time scales and high staffing levels and can be unsuitable for use without adaptation in developing electronic healthcare devices. This paper discusses a set of techniques used in practice to develop software for a particular innovative medical product, an endoscopic camera manipulator. These techniques include identification of potential hazards and tracing their mitigating factors through the project life-cycle.
Keywords :
manipulators; medical computing; medical robotics; safety-critical software; surgery; electronic health care devices; endoscopic camera manipulator; good manufacturing practices; hazardous devices; potential hazard identification; regulatory bodies; safety-related software; surgical robots; Aerospace electronics; Hazards; Injuries; Manufacturing; Medical robotics; Medical services; Robots; Software safety; Standards development; Surges; Marketing of Health Services; Robotics; Software; Surgical Procedures, Operative;