DocumentCode :
128032
Title :
Sensitivity of a 1.5-T MRI system for electromagnetic fields
Author :
Harberts, Dick W. ; van Helvoort, Mark
Author_Institution :
Philips Innovation Services, Electromagn. Competence Center, Eindhoven, Netherlands
fYear :
2014
fDate :
1-4 Sept. 2014
Firstpage :
856
Lastpage :
859
Abstract :
The sensitivity of a 1.5T MRI system has been measured for incident electromagnetic fields. This determines the specification for electronic equipment to be used in an MRI room as well as the specification of MRI room shielding.
Keywords :
biomedical MRI; electromagnetic shielding; MRI room shielding; MRI system; electromagnetic fields; electronic equipment; incident electromagnetic field; Antenna measurements; Electromagnetic compatibility; Magnetic field measurement; Magnetic noise; Magnetic resonance imaging; Signal to noise ratio; EMC; MRI; immunity; shielding; susceptibility;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Electromagnetic Compatibility (EMC Europe), 2014 International Symposium on
Conference_Location :
Gothenburg
Type :
conf
DOI :
10.1109/EMCEurope.2014.6931023