Title :
Sensitivity of a 1.5-T MRI system for electromagnetic fields
Author :
Harberts, Dick W. ; van Helvoort, Mark
Author_Institution :
Philips Innovation Services, Electromagn. Competence Center, Eindhoven, Netherlands
Abstract :
The sensitivity of a 1.5T MRI system has been measured for incident electromagnetic fields. This determines the specification for electronic equipment to be used in an MRI room as well as the specification of MRI room shielding.
Keywords :
biomedical MRI; electromagnetic shielding; MRI room shielding; MRI system; electromagnetic fields; electronic equipment; incident electromagnetic field; Antenna measurements; Electromagnetic compatibility; Magnetic field measurement; Magnetic noise; Magnetic resonance imaging; Signal to noise ratio; EMC; MRI; immunity; shielding; susceptibility;
Conference_Titel :
Electromagnetic Compatibility (EMC Europe), 2014 International Symposium on
Conference_Location :
Gothenburg
DOI :
10.1109/EMCEurope.2014.6931023