Abstract :
A novel electromagnetic characterization of Implantable Cardiac Medical Devices (ICMDs) is presented in this paper. It has been developed in order to determine the immunity level of ICMDs against Radio-Frequency (RF) interference and consists of several Sensitivity Tests, which have been performed in accordance with EMC international standards. The tests have been carried out in an anechoic RF chamber, which guarantees an appropriate management of noise signal injection, shielding against external RF noise sources at the same time. In this way, the operating thresholds of an Implantable Cardiac Defibrillator (ICD) have been achieved at different heartbeat frequency, with and without high-frequency noise. As a result, the comparison between the operating thresholds highlight the electromagnetic susceptibility of the ICD under test.
Keywords :
anechoic chambers (electromagnetic); biomedical measurement; cardiology; defibrillators; electromagnetic compatibility; electromagnetic interference; electromagnetic shielding; immunity testing; prosthetics; EMC characterization; EMC international standards; ICD; ICMD; Implantable Cardiac Defibrillator; anechoic RF chamber; electromagnetic characterization; electromagnetic susceptibility; external RF noise sources; heartbeat frequency; high-frequency noise; immunity level; implantable cardiac medical devices; noise signal injection; operating thresholds; radio-requency interference; sensitivity tests; shielding; Biological system modeling; Electromagnetic compatibility; Electromagnetic interference; Heart beat; Radio frequency; Sensitivity; Standards; Electromagnetic compatibility; Immunity testing; Implantable Cardiac Defibrillator;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Electromagnetic Compatibility (EMC Europe), 2014 International Symposium on
Conference_Location :
Gothenburg
Type :
conf
DOI :
10.1109/EMCEurope.2014.6931026
Filename :
6931026
Link To Document :
بازگشت