• DocumentCode
    1282560
  • Title

    Relaxation in X-Space Magnetic Particle Imaging

  • Author

    Croft, L.R. ; Goodwill, P.W. ; Conolly, S.M.

  • Author_Institution
    Dept. of Bioeng., Univ. of California, Berkeley, Berkeley, CA, USA
  • Volume
    31
  • Issue
    12
  • fYear
    2012
  • Firstpage
    2335
  • Lastpage
    2342
  • Abstract
    Magnetic particle imaging (MPI) is a new imaging modality that noninvasively images the spatial distribution of superparamagnetic iron oxide nanoparticles (SPIOs). MPI has demonstrated high contrast and zero attenuation with depth, and MPI promises superior safety compared to current angiography methods, X-ray, computed tomography, and magnetic resonance imaging angiography. Nanoparticle relaxation can delay the SPIO magnetization, and in this work we investigate the open problem of the role relaxation plays in MPI scanning and its effect on the image. We begin by amending the X-space theory of MPI to include nanoparticle relaxation effects. We then validate the amended theory with experiments from a Berkeley x-space relaxometer and a Berkeley x-space projection MPI scanner. Our theory and experimental data indicate that relaxation reduces SNR and asymmetrically blurs the image in the scanning direction. While relaxation effects can have deleterious effects on the MPI scan, we show theoretically and experimentally that x-space reconstruction remains robust in the presence of relaxation. Furthermore, the role of relaxation in x-space theory provides guidance as we develop methods to minimize relaxation-induced blurring. This will be an important future area of research for the MPI community.
  • Keywords
    X-ray imaging; biomagnetism; image reconstruction; iron compounds; magnetic particles; magnetic relaxation; medical image processing; nanobiotechnology; nanoparticles; Berkeley X-space projection MPI scanner; Berkeley X-space relaxometer; SPIO magnetization; X-space magnetic particle imaging; X-space reconstruction; computed tomography; magnetic resonance imaging angiography; nanoparticle relaxation; relaxation-induced blurring minimization; robust image reconstruction; scanning direction; superparamagnetic iron oxide nanoparticles; zero attenuation; Coils; Convolution; Equations; Image reconstruction; Magnetic resonance imaging; Magnetization; Mathematical model; Ferrofluid relaxation; ferrohydrodynamics; magnetic nanoparticles; magnetic particle imaging (MPI); relaxation; Electromagnetic Fields; Hydrodynamics; Magnetite Nanoparticles; Signal-To-Noise Ratio; Tomography;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.2012.2217979
  • Filename
    6297476