Investigation of the performance of the General Electric ADVANCE positron emission tomograph in 3D mode
Author
Lewellen, T.K. ; Kohlmyer, S.G. ; Miyaoka, R.S. ; Kaplan, M.S. ; Stearns, C.W. ; Schubert, S.F.
Author_Institution
Washington Univ., WA, USA
Volume
43
Issue
4
fYear
1996
fDate
8/1/1996 12:00:00 AM
Firstpage
2199
Lastpage
2206
Abstract
Performance measurements of the General Electric ADVANCE Positron Emission Tomograph operating with the septa retracted (3D mode) were made. All reconstructions were performed with the GE ADVANCE 3D package. Performance tests were carried out with: the NEMA phantoms; a 3D Hoffman phantom; a Data Spectrum torso phantom with lung and cardiac inserts; and the “Utah” 3D evaluation phantom. Data collected included: transaxial and axial resolution, uniformity, recovery coefficients, count rate performance, dead time accuracy, and effect of scatter correction
Keywords
biomedical equipment; positron emission tomography; 3D Hoffman phantom; 3D mode PET; Data Spectrum torso phantom; General Electric ADVANCE positron emission tomograph; NEMA phantoms; Utah 3D evaluation phantom; axial resolution; cardiac inserts; count rate performance; dead time accuracy; lung inserts; medical diagnostic imaging; medical instrumentation; nuclear medicine; recovery coefficients; scatter correction effect; transaxial resolution; uniformity; Collimators; Detectors; Hardware; Image reconstruction; Imaging phantoms; Measurement; Performance evaluation; Radioactive decay; Scattering; Testing;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.531882
Filename
531882
Link To Document