Post-injection transmission scans in a PET camera operating without septa with simultaneous measurement of emission activity contamination
Author :
Smith, Robin J. ; Karp, Joel S.
Author_Institution :
Dept. of Radiol., Pennsylvania Univ., Philadelphia, PA, USA
Volume :
43
Issue :
4
fYear :
1996
fDate :
8/1/1996 12:00:00 AM
Firstpage :
2207
Lastpage :
2212
Abstract :
The authors report on methods developed to reliably perform attenuation correction by post-injection transmission in a volume imaging PET scanner. The method directly measures the emission contamination during the transmission study, using a virtual transmission source position offset by 200 from the actual 68Ge transition source. Events are recorded only if they meet a co-linearity requirement with either the real or virtual source position. The simultaneous measurements of the emission contamination and transmission data remove the need for complex corrections to the emission data and are not subject to activity redistributions between emission and transmission scans performed sequentially. Correction is, necessary, however, for the extra deadtime that varies with the amount of emission activity in the FOV. The extra deadtime is determined from a lookup table of deadtime as a function of detector countrates, which are recorded during the study. The lookup table is based upon phantom measurements. Using patient and phantom data, with both preand post-injection transmission measurements, this method is shown to be reliable for attenuation correction in the body. In addition, it enables the calculation of Standardised Uptake Values for analyzing tumor activity
Keywords :
biomedical equipment; cameras; positron emission tomography; 68Ge transition source; Ge; PET camera operating without septa; activity redistributions; attenuation correction; colinearity requirement; deadtime; emission activity contamination; emission contamination; lookup table; medical diagnostic imaging; nuclear medicine; post-injection transmission scans; tumor activity analysis; virtual transmission source position; Attenuation; Cameras; Contamination; Detectors; Imaging phantoms; Performance evaluation; Pollution measurement; Position measurement; Positron emission tomography; Table lookup;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.531883
Filename :
531883
Link To Document :
بازگشت