Abstract :
The ratio bridge is a common facility in measurement especially in temperature measurement. Four resistances are combined into a resistance network to obtain the ratios that cover almost the range of the ratio bridge. This paper proposes a convenient method to calibrate the linearity of the bridge. It also proposes a new simple way to evaluate the accuracy of the bridge. Finally, a precision thermometry bridge F900, a standard resistance in a thermostatic oil bath and four high stable resistances in a thermostat which made the value of resistances network steady were used to evaluate the new method in National Institute of Metrology.
Keywords :
calibration; electric resistance; oils; temperature measurement; linearity calibration method; precision ratio bridges; precision thermometry bridge F900; standard resistance; temperature measurement; thermostatic oil bath; Conferences; Industrial electronics; Three-dimensional displays; calibration; linearity; precision ratio bridge;