Title :
A simple on-line arterial time-activity curve detector for [O-15] water PET studies
Author :
Wollenweber, S.D. ; Hichwa, R.D. ; Ponto, L.L.B.
Author_Institution :
PET Imaging Center, Iowa Univ., Iowa City, IA, USA
fDate :
6/1/1997 12:00:00 AM
Abstract :
A simple, automated on-line detector system has been fabricated and implemented to detect the arterial time-activity curve (TAG) for bolus-injection [O-15] water PET studies. This system offers two significant improvements over existing systems: a pump mechanism is not required to control arterial blood flow through the detector and dispersion correction of the time-activity curve for dispersion in external tubing is unnecessary. The [O-15] positrons emanating from blood within a thin-walled, 0.134 cm inner-diameter plastic tube are detected by a 0.5 cm wide by 1.0 cm long by 0.1 cm thick plastic scintillator mounted to a miniature PMT. Photon background is reduced to insignificant levels by a 2.0 cm thick cylindrical lead shield. Mean cerebral blood flow (mCBF) determined from an autoradiographic model and from the TAC measured by 1-second automated sampling was compared to that calculated from a TAC acquired using 5-second integrated manual samples. Improvements in timing resolution (1-sec vs. 5-sec) cause small but significant differences between the two sampling methods. Dispersion is minimized due to small tubing diameters, short lengths of tubing between the radial arterial sampling site and the detector and the presence of a 3-way valve 10 cm proximal to the detector
Keywords :
biomedical equipment; blood flow measurement; photomultipliers; positron emission tomography; solid scintillation detectors; water; 0.1 to 2 cm; 1 s; 10 cm; 3-way valve; 5 s; H2O; Pb; [O-15] water PET studies; arterial blood flow; arterial time-activity curve; bolus-injection; dispersion; mean cerebral blood flow; medical diagnostic imaging; miniature PMT; nuclear medicine; photon background; plastic scintillator; pump mechanism; sampling methods; simple on-line arterial time-activity curve detector; thin-walled plastic tube; timing resolution; Automatic control; Blood flow; Control systems; Detectors; Fluid flow measurement; Plastics; Positron emission tomography; Sampling methods; Technical Activities Guide -TAG; Thin wall structures;
Journal_Title :
Nuclear Science, IEEE Transactions on