• DocumentCode
    1292826
  • Title

    The effect of nonlinear signal transformations on bias errors in elastography

  • Author

    Alam, S. Kaisar ; Ophir, J.

  • Author_Institution
    Riverside Res. Inst., New York, NY, USA
  • Volume
    47
  • Issue
    1
  • fYear
    2000
  • Firstpage
    297
  • Lastpage
    303
  • Abstract
    We have reported several artifacts in elastography (1991). These include mechanical artifacts, such as stress concentration, and signal processing artifacts, such as zebras, which are caused by bias errors incurred during the estimation of the peak of correlation functions using a curve-fitting method. We investigate the bias errors and show that bias errors in curve-fitting methods are substantially increased because of nonlinear operations on the echo signals that reduce other errors. We also show that, for typical sampling rates, the bias errors can be ignored in the absence of these nonlinear operations.
  • Keywords
    acoustic signal processing; biomechanics; biomedical ultrasonics; curve fitting; measurement errors; medical signal processing; bias errors; correlation functions; curve-fitting method; echo signals; elastography; mechanical artifacts; nonlinear operations; nonlinear signal transformations; signal processing artifacts; stress concentration; typical sampling rates; zebras; Capacitive sensors; Curve fitting; Delay effects; Delay estimation; Dynamic range; Interpolation; Quantization; Sampling methods; Signal processing; Stress;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/58.818774
  • Filename
    818774