DocumentCode :
1293942
Title :
CO2 laser wavelength standards
Author :
Baird, Kenneth M.
Author_Institution :
National Research Council, Ottawa, Ont., Canada
Issue :
4
fYear :
1976
Firstpage :
445
Lastpage :
447
Abstract :
This paper discusses the possibilities of the CO2 laser for providing a very large number of accurate wavelength-frequency standards in the submillimeter to micrometer region of the spectrum and reviews measurements of the absolute wavelengths, including previously unpublished measurements, leading to a value of 9.305 385 635 μm for the R(14) transition, with an accuracy limited by the present definition of the international meter.
Keywords :
Accuracy; Frequency measurement; Laser stability; Laser transitions; Measurement by laser beam; Standards; Wavelength measurement;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1976.6312260
Filename :
6312260
Link To Document :
بازگشت