• DocumentCode
    1293964
  • Title

    Recent progress in primary Cs beam frequency standards at the PTB

  • Author

    Becker, Gerhard

  • Issue
    4
  • fYear
    1976
  • Firstpage
    458
  • Lastpage
    465
  • Abstract
    The deflection of cesium (Cs) atomic beams using quadrupole and hexapole magnets, or combinations of both, has been investigated. The beam efficiency and the atomic velocity distribution has been calculated for magnets with a cylindrical, conic or parabolic form, respectively. A combination of quadrupole and hexapole magnets allows the selection of atoms in the beam with a relative velocity range of a few percent. A new rectangular frequency modulation method has been developed using modulation widths of multiples of the line width enabling the detection of very small asymmetries of the Cs resonance. By improving the beam optics and lowering the average atomic velocity in the beam to less than 100 m/s, the relative uncertainty of the primary Cs standard CSI of the PTB could be reduced from 1.5 × 10−13 to 1 × 10−13. Measurements of the frequency of the International Atomic Time (TAI) since 1973 showed that on an average the relative frequency deviation from the nominal value is +(10.3 ± 1) × 10−13. TAI frequency measurements with CSI carried out between 1969 and 1973 had revealed a negative TAI-frequency drift.
  • Keywords
    Approximation methods; Atomic beams; Atomic clocks; Atomic measurements; Magnetic flux; Magnetic resonance; Standards;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1976.6312264
  • Filename