DocumentCode
1294066
Title
Comparison of Indiana University Cyclotron Facility Faraday cup proton dosimetry with radiochromic films, a calorimeter, and a calibrated ion chamber
Author
Jones, A.Z. ; Bloch, C.D. ; Hall, E.R. ; Hashemian, R. ; Klein, S.B. ; von Przewoski, B. ; Murray, K.M. ; Foster, C.C.
Author_Institution
Cyclotron Facility, Indiana Univ., Bloomington, IN, USA
Volume
46
Issue
6
fYear
1999
Firstpage
1762
Lastpage
1765
Abstract
The accuracy and utility of the dosimetry system used for radiation effects research with high energy protons at the Indiana University Cyclotron Facility, IUCF, has been confirmed by comparison with an independently calibrated Markus ion chamber, a Schulz water calorimeter and GAFCHROMIC/sup TM/ films.
Keywords
dosimetry; proton detection; radiation monitoring; radiation therapy; Faraday cup proton dosimetry; IUCF; Indiana University Cyclotron Facility; Markus ion chamber; Schulz water calorimeter; dosimetry system; high energy protons; proton beams; proton radiation therapy; radiochromic films; Collimators; Copper; Current measurement; Cyclotrons; Dosimetry; Electrons; Monitoring; Particle beams; Protons; Scattering;