DocumentCode :
1294441
Title :
The effect of noise upon polar-display instantaneous frequency measurement
Author :
Blachman, Nelson M.
Author_Institution :
Western Div., GTE Sylvania, Inc., Mountain View, CA, USA
Issue :
3
fYear :
1976
Firstpage :
214
Lastpage :
221
Abstract :
This system´s rms frequency-measurement error due to noise is calculated, and the effect of a limiter upon IFM performance at high input signal-to-noise ratios (SNR´s) is determined. At high SNR´s, IFM performance with or without a limiter is similar to that of a limiter and discriminator. However, the IFM threshold can be higher (when BT=1) or lower (when BT≫1) than that of a limiter and discriminator, and IFM accuracy exceeds that of the latter at SNR´s below the limiter-discriminator threshold. The IFM threshold is not due to Rice´s `click´ phenomenon but, rather, to second-order noise×noise effects like those in the square-law demodulation of an AM signal, which result in a more gradual decline in performance with decreasing SNR.
Keywords :
display instruments; frequency measurement; noise; IFM performance; IFM threshold; RMS error due to noise; S/R effects; effect of limiters; effect of noise; polar display instantaneous frequency display; Bandwidth; Frequency measurement; Limiting; Low pass filters; Radio frequency; Signal to noise ratio;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1976.6312349
Filename :
6312349
Link To Document :
بازگشت