DocumentCode :
1294689
Title :
Frequency dependence of 50-Ω coaxial open-circuit reflection standard
Author :
Dibeneditto, J. ; Uhlir, A.
Author_Institution :
US Air Force, RADC, Hanscom AFB, USA
Issue :
3
fYear :
1981
Firstpage :
228
Lastpage :
229
Abstract :
The effective position of the open circuit in precision 7-mm 50-Ω coaxial transmission line is found experimentally to be located beyond the physical end of the center conductor by a distance 1.198+2.69×10-10f2 millimeters at f megahertz (<;18000). Because this determination is substantially identical to published theoretical values, the open circuit is recommended as a calibration standard for network analyzers.
Keywords :
calibration; coaxial cables; measurement standards; network analysers; calibration; coaxial transmission line; frequency dependence; network analyzers; open circuit reflection standard; Calibration; Capacitance; Fluctuations; Frequency dependence; Masers; Reflection; Standards;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1981.6312385
Filename :
6312385
Link To Document :