    Frequency dependence of 50-Ω coaxial open-circuit reflection standard

  • Author

    Dibeneditto, J. ; Uhlir, A.

  • Author_Institution
    US Air Force, RADC, Hanscom AFB, USA
  • Issue
    3
  • fYear
    1981
  • Firstpage
    228
  • Lastpage
    229
  • Abstract
    The effective position of the open circuit in precision 7-mm 50-Ω coaxial transmission line is found experimentally to be located beyond the physical end of the center conductor by a distance 1.198+2.69×10-10f2 millimeters at f megahertz (<;18000). Because this determination is substantially identical to published theoretical values, the open circuit is recommended as a calibration standard for network analyzers.
  • Keywords
    calibration; coaxial cables; measurement standards; network analysers; calibration; coaxial transmission line; frequency dependence; network analyzers; open circuit reflection standard; Calibration; Capacitance; Fluctuations; Frequency dependence; Masers; Reflection; Standards;