• DocumentCode
    1294961
  • Title

    A triaxial accelerometer and portable data processing unit for the assessment of daily physical activity

  • Author

    Bouten, Carlijn V C ; Koekkoek, Karel T M ; Verduin, Maarten ; Kodde, Rens ; Janssen, Jan D.

  • Author_Institution
    Div. of Comput. & Exp. Mech., Univ. of Technol., Eindhoven, Netherlands
  • Volume
    44
  • Issue
    3
  • fYear
    1997
  • fDate
    3/1/1997 12:00:00 AM
  • Firstpage
    136
  • Lastpage
    147
  • Abstract
    The present study describes the development of a triaxial accelerometer (TA) and a portable data processing unit for the assessment of daily physical activity. The TA is composed of three orthogonally mounted uniaxial piezoresistive accelerometers and can be used to register accelerations covering the amplitude and frequency ranges of human body acceleration. Interinstrument and test-retest experiments showed that the offset and the sensitivity of the TA were equal for each measurement direction and remained constant on two measurement days. Transverse sensitivity was significantly different for each measurement direction, but did not influence accelerometer output (<3% of the sensitivity along the main axis). The data unit enables the on-line processing of accelerometer output to a reliable estimator of physical activity over eight-day periods. Preliminary evaluation of the system in 13 male subjects during standardized activities in the laboratory demonstrated a significant relationship between accelerometer output and energy expenditure due to physical activity, the standard reference for physical activity (r=0.89). Shortcomings of the system are its low sensitivity to sedentary activities and the inability to register static exercise. The validity of the system for the assessment of normal daily physical activity and specific activities outside the laboratory should be studied in free-living subjects.
  • Keywords
    accelerometers; biological techniques; biomechanics; 8 d; daily physical activity assessment; energy expenditure; free-living subjects; human body acceleration; laboratory; male subjects; measurement direction; orthogonally mounted uniaxial piezoresistive accelerometers; portable data processing unit; sedentary activities; standardized activities; static exercise; test-retest experiments; transverse sensitivity; triaxial accelerometer; Acceleration; Accelerometers; Data processing; Frequency; Humans; Laboratories; Mechanical sensors; Piezoresistance; Sensor phenomena and characterization; Testing; Acceleration; Activities of Daily Living; Automatic Data Processing; Calibration; Energy Metabolism; Equipment Design; Evaluation Studies as Topic; Exercise; Humans; Male; Monitoring, Physiologic; Movement; Posture; Sensitivity and Specificity;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.554760
  • Filename
    554760