Title :
Equivalent circuit of an open-ended coaxial line in a lossy dielectric
Author :
Stuchly, M.A. ; Stuchly, S.S.
Author_Institution :
Radiation Protection Bur., Health & Welfare, Ottawa, Ont., Canada
fDate :
6/1/1982 12:00:00 AM
Abstract :
An open-ended coaxial line was used as a sensor in measurement of the permittivity of lossy dielectrics at radio and microwave frequencies. A simplified equivalent circuit related the permittivity to the measured input reflected coefficient. A more complete equivalent circuit is proposed and errors in the permittivity measurements resulting from the simplification are analyzed.
Keywords :
coaxial cables; electric sensing devices; equivalent circuits; microwave measurement; permittivity measurement; electric sensing devices; errors; input reflected coefficient; lossy dielectric; microwave frequencies; open-ended coaxial line; permittivity measurement; radio frequencies; sensor; Equivalent circuits; Frequency measurement; Noise; Permittivity; Permittivity measurement; Transmission line measurements;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :