Title :
A 4-GHz band low-noise measurement system
Author :
Kato, Yoshihiko ; Yokoshima, Ichiro
Author_Institution :
Electrotech. Lab., Ibaraki, Japan
fDate :
3/1/1987 12:00:00 AM
Abstract :
A 4-GHz-band noise measurement system has been built for calibrating microwave low-noise sources accurately. The system reduces the contribution of several errors occurring in conventional systems. The system is composed of a highly stable and sensitive Dicke-type comparison radiometer, whose front end is cooled with liquid nitrogen, and two highly stable standard noise sources: one using liquid nitrogen and the other operating at the ice point. It is also computerized to provide automatic measurement operation and real-time data processing. The ultimate calibration error is within about ±0.4 K, for all temperatures up to 350 K, and is within ±0.2 K near 80 K and 273 K.
Keywords :
computerised instrumentation; electric noise measurement; microwave measurement; radiometers; 4 GHz; Dicke comparison radiometer; automatic measurement; computerised instrumentation; liquid N2; microwave low-noise sources; real-time data processing; standard noise sources; Calibration; Microwave measurements; Microwave radiometry; Noise; Noise measurement; Standards; Temperature measurement;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :
10.1109/TIM.1987.6312631