• DocumentCode
    1296394
  • Title

    Twenty years of SI ohm determinations at NML

  • Author

    Small, Greig W.

  • Author_Institution
    National Measurement Laboratory, CSIRO Division of Applied Physics, Lindheld, NSW, Australia 2070
  • Issue
    2
  • fYear
    1987
  • fDate
    6/1/1987 12:00:00 AM
  • Firstpage
    190
  • Lastpage
    195
  • Abstract
    The National Measurement Laboratory (NML) of Australia has maintained an absolute standard of resistance based on a calculable capacitor for more than 20 years. Since its inception, various developments have been incorporated which contribute to a reduction in the overall uncertainty. A reassessment of uncertainties gives a figure now of 6.2 in 108 for the ohm. A systematic error, introduced in 1974, has been identified, and a corrrection of −0.09 ppm should be made to resistances expressed in terms of the NML ohm since that time.
  • Keywords
    Bridge circuits; Capacitors; Electrical resistance measurement; Resistance; Resistors; Standards; Uncertainty;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1987.6312666
  • Filename