DocumentCode :
1296574
Title :
Methods adopted at LCIE for solving basic problems in the determination of AC-DC transfer differences of thermal converters
Author :
Erard, Luc ; Knosp, Robert ; Soucek, Radomir
Author_Institution :
Laboratoire Central des Industries Electriques, Fontenay-aux-Roses, France
Issue :
2
fYear :
1987
fDate :
6/1/1987 12:00:00 AM
Firstpage :
337
Lastpage :
340
Abstract :
The absolute determination of the ac-dc transfer differences by a step-up method involves some problems such as the knowledge of the ac-dc transfer differences öf the reference element and the determination of the error arising from the use of thermoelements at full range and fractional range during the step-up procedure. The methods used to take care of these errors are described.
Keywords :
Accuracy; Frequency conversion; Frequency measurement; Generators; Resistors; Uncertainty; Voltage measurement;