• DocumentCode
    1298408
  • Title

    Coil design for real and sham transcranial magnetic stimulation

  • Author

    Ruohonen, Jarmo ; Ollikainen, Marko ; Nikouline, Vadim ; Virtanen, Juha ; Ilmoniemi, Risto J.

  • Author_Institution
    Central Hosp., Helsinki Univ., Finland
  • Volume
    47
  • Issue
    2
  • fYear
    2000
  • Firstpage
    145
  • Lastpage
    148
  • Abstract
    Transcranial magnetic stimulation (TMS) can be used to excite the human cortex noninvasively. TMS also activates scalp muscles and sensory receptors; additionally, the loud sound from the stimulating coil activates auditory pathways. These side effects complicate the interpretation of the results of TMS studies. For control experiments, the authors have designed a coil that can produce both real and sham stimulation without moving the coil. The sham TMS is similar to the real TMS, except for the different relative direction of the currents in the two loops of the figure of eight coil. While the real TMS elicited activation of hand muscles, sham TMS had no such effect; however, the auditory-evoked potentials were similar.
  • Keywords
    auditory evoked potentials; biomagnetism; biomedical equipment; brain; coils; electroencephalography; EEG; currents relative direction; figure of eight coil; hand muscles activation; medical instrumentation; real transcranial magnetic stimulation; sham transcranial magnetic stimulation; side effects; Biomedical engineering; Cerebral cortex; Coils; Electroencephalography; Hospitals; Humans; Magnetic heads; Magnetic stimulation; Muscles; Scalp; Electroencephalography; Electromyography; Equipment Design; Evoked Potentials, Auditory; Humans; Muscle, Skeletal; Physical Stimulation; Reference Values; Transcranial Magnetic Stimulation;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.821731
  • Filename
    821731