• DocumentCode
    1312131
  • Title

    Radiowave propagation from a tissue-implanted source at 418 MHz and 916.5 MHz

  • Author

    Scanlon, William G. ; Burns, J. Brian ; Evans, Noel E.

  • Author_Institution
    Sch. of Electr. & Mech, Eng., Ulster Univ., Newtownabbey, UK
  • Volume
    47
  • Issue
    4
  • fYear
    2000
  • fDate
    4/1/2000 12:00:00 AM
  • Firstpage
    527
  • Lastpage
    534
  • Abstract
    Tissue-implanted ultra-high frequency (UHF) radio devices are being employed in both humans and animals for telemetry and telecommand applications. This paper describes the experimental measurement and electromagnetic modeling of propagation from 418-MHz and 916.5-MHz sources placed in the human vagina. Whole-body homogeneous and semi-segmented software models were constructed using data from the Visible Human Project. Bodyworn radiation efficiencies for a vaginally placed 418-MHz source were calculated using finite-difference time-domain and ranged between 1.6% and 3.4% (corresponding to net body losses of between 14.7 and 18.0 dB). Greater losses were encountered at 916.5 MHz, with efficiencies between 0.36% and 0.46% (net body loss ranging between 23.4 and 24.4 dB). Practical measurements mere in good agreement with simulations, to within 2 dB at 418 MHz and 3 dB at 916.5 MHz. The degree of tissue-segmentation for whole-body models was found to have a minimal effect on calculated azimuthal radiation patterns and bodyworn radiation efficiency, provided the region surrounding the implanted source was sufficiently detailed.
  • Keywords
    UHF radio propagation; antenna radiation patterns; biological tissues; biomedical telemetry; finite difference time-domain analysis; patient monitoring; prosthetics; radiotelemetry; 14.7 to 18 dB; 23.4 to 24.4 dB; 418 MHz; 916.5 MHz; FDTD calculation; UHF radio devices; Visible Human Project; azimuthal radiation patterns; bodyworn radiation efficiencies; electromagnetic modeling; human vagina; radiowave propagation; semi-segmented software models; telecommand; telemetry; tissue-implanted source; tissue-segmentation degree; whole-body homogeneous models; Animals; Application software; Electromagnetic measurements; Electromagnetic modeling; Frequency; Humans; Radiowave propagation; Telecommunications; Telemetry; UHF measurements; Algorithms; Anatomy, Cross-Sectional; Electric Conductivity; Electromagnetics; Female; Humans; Middle Aged; Models, Biological; Physical Stimulation; Prostheses and Implants; Radio Waves; Vagina;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.828152
  • Filename
    828152