Title :
Ventricular volume measurement from a multiplanar transesophageal ultrasonic imaging system: an in vitro study
Author :
Martin, Roy W. ; Bashein, Gerard ; Detmer, Paul R. ; Moritz, William E.
Author_Institution :
Washington Univ., Seattle, WA, USA
fDate :
5/1/1990 12:00:00 AM
Abstract :
A system to assess the feasibility of using multiple transesophageal ultrasonic images to measure left-ventricular volume is discussed. The system includes a special transesophageal probe with a micromanipulator for acquiring cardiac images in multiple planes with known interplanar spatial relationship and an offline processing system to compute the volume. In vitro studies with the probe demonstrated that the distance between two targets in space can be identified within 2 mm ( sigma =0.4 mm) for points in the imaging plane and 3.4 mm ( sigma =0.5 mm) for points not lying in the imaging plane. This gives an average accuracy of +or-6.5% for distances greater than 4.5 cm. Comparison of ultrasonic measurements of the volume of water-filled balloons and excised hearts to the volume required to fill them revealed a correlation coefficient of 0.992, a regression line with a slope of 1.0 and an ordinate intercept at 0.2 mL, and a standard error of the estimate of 8 mL.
Keywords :
acoustic imaging; biomedical measurement; biomedical ultrasonics; cardiology; ultrasonic measurement; volume measurement; 2 mm; 3.4 mm; excised hearts; interplanar spatial relationship; left ventricular volume measurement; medical ultrasound; micromanipulator; multiplanar transesophageal ultrasonic imaging system; offline processing system; ordinate intercept; regression line; transesophageal probe; water-filled balloons; Biomedical measurements; Catheters; Heart; Patient monitoring; Pressure measurement; Probes; Surgery; Ultrasonic imaging; Ultrasonic variables measurement; Volume measurement; Calibration; Esophagus; Humans; Image Processing, Computer-Assisted; Stroke Volume; Ultrasonography;
Journal_Title :
Biomedical Engineering, IEEE Transactions on