DocumentCode
1316352
Title
A new technique for line interference monitoring and reduction in biopotential amplifiers
Author
Ider, Yusuf Ziya ; Köymen, Hayrettin
Author_Institution
Dept. of Electr. & Electron. Eng., Middle East Tech. Univ., Ankara, Turkey
Volume
37
Issue
6
fYear
1990
fDate
6/1/1990 12:00:00 AM
Firstpage
624
Lastpage
631
Abstract
Hardware developed to record the common mode line frequency signal on the body simultaneously with the ECG lead signals of a 15-channel computerized cardiograph is described. This interference reference signal and its quadrature, obtained by software, are linearly combined to be subtracted from any one of the channels to reduce the interference to below the quantization level of the 12 b A/D converter. Coefficients of the linear combination are estimated using linear regression, which is applied to the relatively isoelectric regions of the data, excluding the QRS complexes. Since the interference reference signal is available in real time, simultaneously with the ECG signals, another software approach is adopted in which an adaptive interference reduction algorithm is used to cope with varying interference. A recursive-least-squares algorithm with forgetting factor is used to update the coefficients. This updating mechanism is gated by the output of a software QRS detector. Results regarding the performance of both the offline and the adaptive algorithms are given, and the effects of nonisoelectric portions of the ECG lead signals on the estimation of the coefficients are quantified.
Keywords
bioelectric potentials; computerised monitoring; computerised signal processing; electrocardiography; interference (signal); interference suppression; medical computing; 15-channel computerized cardiograph; A/D converter; ECG lead signals; QRS complexes; biopotential amplifiers; common mode line frequency signal; forgetting factor; interference reference signal; line interference monitoring; line interference reduction; linear regression; quadrature; recursive-least-squares algorithm; software QRS detector; Cardiography; Computerized monitoring; Detectors; Electrocardiography; Frequency; Hardware; Interference; Linear regression; Quantization; Software algorithms; Algorithms; Amplifiers; Electrocardiography; Microcomputers; Reference Standards; Signal Processing, Computer-Assisted; Software;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/10.55666
Filename
55666
Link To Document