Title :
Electromagnetic Interference of Cardiac Rhythmic Monitoring Devices to Radio Frequency Identification: Analytical Analysis and Mitigation Methodology
Author :
Ogirala, Ajay ; Stachel, Joshua R. ; Mickle, Marlin H.
Author_Institution :
Dept. of Electr. & Comput. Eng., Univ. of Pittsburgh, Pittsburgh, PA, USA
Abstract :
Increasing density of wireless communication and development of radio frequency identification (RFID) technology in particular have increased the susceptibility of patients equipped with cardiac rhythmic monitoring devices (CRMD) to environmental electro magnetic interference (EMI). Several organizations reported observing CRMD EMI from different sources. This paper focuses on mathematically analyzing the energy as perceived by the implanted device, i.e., voltage. Radio frequency (RF) energy transmitted by RFID interrogators is considered as an example. A simplified front-end equivalent circuit of a CRMD sensing circuitry is proposed for the analysis following extensive black-box testing of several commercial pacemakers and implantable defibrillators. After careful understanding of the mechanics of the CRMD signal processing in identifying the QRS complex of the heart-beat, a mitigation technique is proposed. The mitigation methodology introduced in this paper is logical in approach, simple to implement and is therefore applicable to all wireless communication protocols.
Keywords :
biomedical equipment; defibrillators; electrocardiography; electromagnetic interference; equivalent circuits; medical signal processing; pacemakers; patient monitoring; radiofrequency identification; CRMD sensing circuitry; QRS complex; RFID interrogators; analytical analysis; blackbox testing; cardiac rhythmic monitoring devices; electromagnetic interference; frontend equivalent circuit; heartbeat; implantable defibrillators; implanted device; mitigation methodology; pacemakers; radiofrequency identification; radiofrequency identification technology; signal processing; wireless communication; Biomedical monitoring; Cardiology; Defibrillators; Heart; Pacemakers; Radiofrequency identification; Wireless communication; Cardiac rhythmic monitoring devices (CRMD); QRS complex; electro magnetic interference (EMI); electro magnetic interference mitigation; heart signal; implantable cardiac defibrillators (ICD); pacemaker; radio frequency identification (RFID); Arrhythmias, Cardiac; Defibrillators, Implantable; Electromagnetic Fields; Electromagnetic Phenomena; Equipment Failure Analysis; Equipment Safety; Humans; Pacemaker, Artificial; Radio Frequency Identification Device; Radio Waves;
Journal_Title :
Information Technology in Biomedicine, IEEE Transactions on
DOI :
10.1109/TITB.2011.2163640