• DocumentCode
    1320229
  • Title

    Electromagnetic Interference of Cardiac Rhythmic Monitoring Devices to Radio Frequency Identification: Analytical Analysis and Mitigation Methodology

  • Author

    Ogirala, Ajay ; Stachel, Joshua R. ; Mickle, Marlin H.

  • Author_Institution
    Dept. of Electr. & Comput. Eng., Univ. of Pittsburgh, Pittsburgh, PA, USA
  • Volume
    15
  • Issue
    6
  • fYear
    2011
  • Firstpage
    848
  • Lastpage
    853
  • Abstract
    Increasing density of wireless communication and development of radio frequency identification (RFID) technology in particular have increased the susceptibility of patients equipped with cardiac rhythmic monitoring devices (CRMD) to environmental electro magnetic interference (EMI). Several organizations reported observing CRMD EMI from different sources. This paper focuses on mathematically analyzing the energy as perceived by the implanted device, i.e., voltage. Radio frequency (RF) energy transmitted by RFID interrogators is considered as an example. A simplified front-end equivalent circuit of a CRMD sensing circuitry is proposed for the analysis following extensive black-box testing of several commercial pacemakers and implantable defibrillators. After careful understanding of the mechanics of the CRMD signal processing in identifying the QRS complex of the heart-beat, a mitigation technique is proposed. The mitigation methodology introduced in this paper is logical in approach, simple to implement and is therefore applicable to all wireless communication protocols.
  • Keywords
    biomedical equipment; defibrillators; electrocardiography; electromagnetic interference; equivalent circuits; medical signal processing; pacemakers; patient monitoring; radiofrequency identification; CRMD sensing circuitry; QRS complex; RFID interrogators; analytical analysis; blackbox testing; cardiac rhythmic monitoring devices; electromagnetic interference; frontend equivalent circuit; heartbeat; implantable defibrillators; implanted device; mitigation methodology; pacemakers; radiofrequency identification; radiofrequency identification technology; signal processing; wireless communication; Biomedical monitoring; Cardiology; Defibrillators; Heart; Pacemakers; Radiofrequency identification; Wireless communication; Cardiac rhythmic monitoring devices (CRMD); QRS complex; electro magnetic interference (EMI); electro magnetic interference mitigation; heart signal; implantable cardiac defibrillators (ICD); pacemaker; radio frequency identification (RFID); Arrhythmias, Cardiac; Defibrillators, Implantable; Electromagnetic Fields; Electromagnetic Phenomena; Equipment Failure Analysis; Equipment Safety; Humans; Pacemaker, Artificial; Radio Frequency Identification Device; Radio Waves;
  • fLanguage
    English
  • Journal_Title
    Information Technology in Biomedicine, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1089-7771
  • Type

    jour

  • DOI
    10.1109/TITB.2011.2163640
  • Filename
    6018306