• DocumentCode
    1322853
  • Title

    Evaluation of Harmonic Signals for the Detection of Magnetic Nanoparticles

  • Author

    Yoshida, T. ; Othman, N.B. ; Tsubaki, T. ; Takamiya, J. ; Enpuku, K.

  • Author_Institution
    Dept. of Electr. & Electron. Eng., Kyushu Univ., Fukuoka, Japan
  • Volume
    48
  • Issue
    11
  • fYear
    2012
  • Firstpage
    3788
  • Lastpage
    3791
  • Abstract
    We have evaluated the use of harmonic signals for detecting magnetic nanoparticles (MNPs) in biomedical applications. We studied the relationship between the harmonic signals and the distribution of magnetic moments in MNP samples. We measured the harmonic signals and frequency dependence of the ac susceptibility along with static M-H curves. These experimental results were analyzed to obtain the magnetic moment distribution in real samples. We were able to clarify what proportion of MNPs contributes to the harmonic signal. MNPs could be classified according to three types. Type-1 contributes to the fundamental signal only. Type-2 contributes to both the fundamental and harmonic signals. Finally, Type-3 does not contribute to any signals. It is important to choose a sample having a large portion of Type-2 MNPs for the harmonic signal-based detection.
  • Keywords
    biomedical materials; magnetic moments; magnetic particles; magnetic susceptibility; medical signal detection; medical signal processing; nanoparticles; MNP; ac susceptibility; harmonic signal detection; magnetic moments; magnetic nanoparticles; signal classification; Harmonic analysis; Magnetic moments; Magnetic resonance imaging; Magnetic susceptibility; Nanoparticles; Suspensions; Bionanotechnology; medical diagnostic imaging; nanoparticles; nonlinear magnetics;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/TMAG.2012.2198198
  • Filename
    6333021