DocumentCode :
1323012
Title :
Maintenance of a Laboratory Unit of Voltage
Author :
Dunn, Andrew F.
Author_Institution :
Division of Physics, National Research Council, Ottawa, Ont., Canada.
Issue :
1
fYear :
1971
Firstpage :
2
Lastpage :
10
Abstract :
Many of the problems involved in maintaining a laboratory reference group of standard cells with a reliability better than one part per million are outlined and solutions suggested. Methods are presented for reducing the number of measurements by approximately 40 percent from the ideal without seriously reducing the accuracy achieved, and procedures for assessing the quality of the group are indicated. Suggestions are given for preferred methods to be used for groups of 4, 8, or 12 cells, and as an example of these procedures the 20-month history of a new group of standard cells is illustrated.
Keywords :
Detectors; Equations; Laboratories; Mathematical model; Potentiometers; Temperature measurement; Voltage measurement;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1971.5570731
Filename :
5570731
Link To Document :
بازگشت