Title :
Image Reconstruction in Microwave Tomography Using a Dielectric Debye Model
Author :
Fhager, Andreas ; Gustafsson, Mats ; Nordebo, Sven
Author_Institution :
Dept. of Signal & Syst., Chalmers Univ. of Technol., Goteborg, Sweden
Abstract :
In this paper, quantitative dielectric image reconstruction based on broadband microwave measurements is investigated. A time-domain-based algorithm is derived where Debye model parameters are reconstructed in order to take into account the strong dispersive behavior found in biological tissue. The algorithm is tested with experimental and numerical data in order to verify the algorithm and to investigate improvements in the reconstructed image resulting from the improved description of the dielectric properties of the tissue when using broadband data. The comparison is made in relation to the more commonly used conductivity model. For the evaluation, two examples were considered, the first was a lossy saline solution and the second was less lossy tap water. Both liquids are strongly dispersive and used as a background medium in the imaging examples. The results show that the Debye model algorithm is of most importance in the tap water for a bandwidth of more than 1.5 GHz. Also the saline solution exhibits a dispersive behavior but since the losses restrict the useful bandwidth, the Debye model is of less significance even if somewhat larger and stronger artifacts can be seen in the conductivity model reconstructions.
Keywords :
bioelectric phenomena; biological tissues; dielectric properties; image reconstruction; medical image processing; microwave imaging; tomography; Debye model algorithm; biological tissue; broadband microwave measurement; conductivity model; dielectric Debye model; lossy saline solution; lossy tap water; microwave tomography; quantitative dielectric image reconstruction; time-domain-based algorithm; tissue dielectric property; Antenna measurements; Biomedical measurements; Conductivity; Data models; Dispersion; Image reconstruction; Time domain analysis; Debye model; dispersion; finite-difference time-domain (FDTD) methods; microwave imaging; microwave measurements; Computer Simulation; Humans; Image Interpretation, Computer-Assisted; Imaging, Three-Dimensional; Microwaves; Models, Biological; Plethysmography, Impedance; Tomography;
Journal_Title :
Biomedical Engineering, IEEE Transactions on
DOI :
10.1109/TBME.2011.2168606