DocumentCode
1331905
Title
The Improvement of the Noninvasive Power-Supply System Using Magnetic Coupling for Medical Implants
Author
Suzuki, Satoshi ; Ishihara, Manabu ; Kobayashi, Yoshiyuki
Author_Institution
Dept. of Electr. & Comput. Eng., Oyama Nat. Coll. of Technol., Tochigi, Japan
Volume
47
Issue
10
fYear
2011
Firstpage
2811
Lastpage
2814
Abstract
We proposed a novel noninvasive power-supply system for medical implants at a deep position in the body. The system consists of the magnetic coupling and the small generator. The former system could keep the power of 1 W supplying until 2.5 cm from the skin and the maximum output power was 3.1 W at 1.5 cm. We have manufactured more powerful generators as the next step. The improvement points are to decrease or eliminate the interference between the magnets at the implanted rotor. As a result, the output power increases 6.6 W and the 1 W supplying distance is extended to 3.0 cm.
Keywords
electric generators; permanent magnets; prosthetic power supplies; skin; distance 1.5 cm; distance 3.0 cm; magnetic coupling; medical implants; noninvasive power-supply system; power 3.1 W; power 6.6 W; skin; Amorphous magnetic materials; Couplings; Generators; Interference; Magnetic flux; Power generation; Rotors; Magnetic coupling; medical implant; noninvasive; permanent magnet; power supply;
fLanguage
English
Journal_Title
Magnetics, IEEE Transactions on
Publisher
ieee
ISSN
0018-9464
Type
jour
DOI
10.1109/TMAG.2011.2151273
Filename
6028144