• DocumentCode
    1331905
  • Title

    The Improvement of the Noninvasive Power-Supply System Using Magnetic Coupling for Medical Implants

  • Author

    Suzuki, Satoshi ; Ishihara, Manabu ; Kobayashi, Yoshiyuki

  • Author_Institution
    Dept. of Electr. & Comput. Eng., Oyama Nat. Coll. of Technol., Tochigi, Japan
  • Volume
    47
  • Issue
    10
  • fYear
    2011
  • Firstpage
    2811
  • Lastpage
    2814
  • Abstract
    We proposed a novel noninvasive power-supply system for medical implants at a deep position in the body. The system consists of the magnetic coupling and the small generator. The former system could keep the power of 1 W supplying until 2.5 cm from the skin and the maximum output power was 3.1 W at 1.5 cm. We have manufactured more powerful generators as the next step. The improvement points are to decrease or eliminate the interference between the magnets at the implanted rotor. As a result, the output power increases 6.6 W and the 1 W supplying distance is extended to 3.0 cm.
  • Keywords
    electric generators; permanent magnets; prosthetic power supplies; skin; distance 1.5 cm; distance 3.0 cm; magnetic coupling; medical implants; noninvasive power-supply system; power 3.1 W; power 6.6 W; skin; Amorphous magnetic materials; Couplings; Generators; Interference; Magnetic flux; Power generation; Rotors; Magnetic coupling; medical implant; noninvasive; permanent magnet; power supply;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/TMAG.2011.2151273
  • Filename
    6028144