DocumentCode
1334801
Title
High-density flexible interconnect for two-dimensional ultrasound arrays
Author
Fiering, Jason O. ; Hultman, Peter ; Lee, Warren ; Light, Edward D. ; Smith, Stephen W.
Author_Institution
Dept. of Electr. & Comput. Eng., North Carolina State Univ., Raleigh, NC, USA
Volume
47
Issue
3
fYear
2000
fDate
5/1/2000 12:00:00 AM
Firstpage
764
Lastpage
770
Abstract
We present a method for fabricating flexible multilayer circuits for interconnection to 2-D array ultrasound transducers. In addition, we describe four 2-D arrays in which such flexible interconnect is implemented, including transthoracic arrays with 438 channels operating at up to 7 MHz and intracardiac catheter arrays with 70 channels operating at up to 7 MHz. We employ thin and thick film microfabrication techniques to batch produce the interconnect circuits with minimum dimensions of 12-/spl mu/m lines, 40-/spl mu/m vias, and 150-/spl mu/m array pitch. The arrays show 50-/spl Omega/ insertion loss of -60 to -84 dB and a fractional bandwidth of 27 to 67%. The arrays are used to obtain real time, in vivo volumetric scans.
Keywords
biomedical ultrasonics; interconnections; ultrasonic transducer arrays; -60 to -84 dB; 7 MHz; batch production; high-density interconnect; intracardiac catheter array; medical imaging; multilayer flexible circuit; real-time in-vivo volumetric scan; thick film microfabrication; thin film microfabrication; transthoracic array; two-dimensional ultrasound transducer array; Catheters; Flexible printed circuits; Insertion loss; Integrated circuit interconnections; Nonhomogeneous media; Thick film circuits; Thick films; Ultrasonic imaging; Ultrasonic transducer arrays; Ultrasonic transducers;
fLanguage
English
Journal_Title
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher
ieee
ISSN
0885-3010
Type
jour
DOI
10.1109/58.842067
Filename
842067
Link To Document