• DocumentCode
    1335116
  • Title

    A Robotic Cadaveric Gait Simulator With Fuzzy Logic Vertical Ground Reaction Force Control

  • Author

    Aubin, Patrick M. ; Whittaker, Eric ; Ledoux, William R.

  • Author_Institution
    Dept. of Electr. Eng., Univ. of Washington, Seattle, WA, USA
  • Volume
    28
  • Issue
    1
  • fYear
    2012
  • Firstpage
    246
  • Lastpage
    255
  • Abstract
    Lower limb dynamic cadaveric gait simulators are useful to investigate the biomechanics of the foot and ankle, but many systems have several common limitations, which include simplified tendon forces, nonphysiologic tibial kinematics, greatly reduced velocities, scaled body weight (BW), and, most importantly, trial-and-error vertical ground reaction force (vGRF) control. This paper presents the design, development, and validation of the robotic gait simulator (RGS), which addresses these limitations. A 6-degrees-of-freedom (6-DOF) parallel robot was utilized as part of the RGS to recreate the relative tibia to ground motion. A custom-designed nine-axis proportional-integral-derivative (PID) force-control tendon actuation system provided force to the extrinsic tendons of the cadaveric lower limb. A fuzzy logic vGRF controller was developed, which altered tendon forces in real time and iteratively adjusted the robotic trajectory in order to track a target vGRF. The RGS was able to accurately reproduce 6-DOF tibial kinematics, tendon forces, and vGRF with a cadaveric lower limb. The fuzzy logic vGRF controller was able to track the target in vivo vGRF with an average root-mean-square error of only 5.6% BW during a biomechanically realistic 3/4 BW, 2.7-s stance phase simulation.
  • Keywords
    biomechanics; force control; fuzzy control; medical robotics; motion control; robot dynamics; robot kinematics; three-term control; 6-DOF tibial kinematics; 6-degrees-of-freedom parallel robot; ankle biomechanics; foot biomechanics; fuzzy logic vGRF controller; fuzzy logic vertical ground reaction force control; lower limb dynamic cadaveric gait simulator; nine-axis proportional-integral-derivative force-control tendon actuation system; nonphysiologic tibial kinematics; robotic cadaveric gait simulator; robotic trajectory; root-mean-square error; scaled body weight; tendon force; trial-and-error vertical ground reaction force; Foot; Force; Fuzzy logic; Fuzzy sets; Robots; Tendons; Trajectory; Force control; gait simulation; medical robots and systems; neural and fuzzy control; parallel robotics;
  • fLanguage
    English
  • Journal_Title
    Robotics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1552-3098
  • Type

    jour

  • DOI
    10.1109/TRO.2011.2164958
  • Filename
    6029994