    Temperature Dependent Operation of PSAPD-Based Compact Gamma Camera for SPECT Imaging

  • Author

    Kim, Sangtaek ; McClish, Mickel ; Alhassen, Fares ; Seo, Youngho ; Shah, Kanai S. ; Gould, Robert G.

  • Author_Institution
    Phys. Res. Lab., Univ. of California, San Francisco, CA, USA
  • Volume
    58
  • Issue
    5
  • fYear
    2011
  • Firstpage
    2169
  • Lastpage
    2174
  • Abstract
    We investigated the dependence of image quality on the temperature of a position sensitive avalanche photodiode (PSAPD)-based small animal single photon emission computed tomography (SPECT) gamma camera with a CsI:Tl scintillator. Currently, nitrogen gas cooling is preferred to operate PSAPDs in order to minimize the dark current shot noise. Being able to operate a PSAPD at a relatively high temperature (e.g., 5°C) would allow a more compact and simple cooling system for the PSAPD. In our investigation, the temperature of the PSAPD was controlled by varying the flow of cold nitrogen gas through the PSAPD module and varied from -40°C to 20°C. Three experiments were performed to demonstrate the performance variation over this temperature range. The point spread function (PSF) of the gamma camera was measured at various temperatures, showing variation of full-width-half-maximum (FWHM) of the PSF. In addition, a 99 mTc-pertechnetate (140 keV) flood source was imaged and the visibility of the scintillator segmentation (16 × 16 array, 8 mm × 8 mm area, 400 μ m pixel size) at different temperatures was evaluated. Comparison of image quality was made at -25°C and 5°C using a mouse heart phantom filled with an aqueous solution of 99 m Tc-pertechnetate and imaged using a 0.5 mm pinhole collimator made of tungsten. The reconstructed image quality of the mouse heart phantom at 5°C degraded in comparision to the reconstructed image quality at -25°C. However, the defect and structure of the mouse heart phantom were clearly observed, showing the feasibility of operating PSAPDs for SPECT imaging at 5°C, a temperature that would not need the nitrogen cooling. All PSAPD evaluations were conducted with an applied bias voltage that allowed the highest gain at a given temperature.
  • Keywords
    avalanche photodiodes; image reconstruction; medical image processing; phantoms; single photon emission computed tomography; PSAPD-based compact gamma camera; SPECT imaging; dark current shot noise; image quality; image reconstruction; mouse heart phantom; point spread function; position sensitive avalanche photodiode; scintillator segmentation; small animal single photon emission computed tomography; temperature 20 degC to 40 degC; temperature 273 K; temperature 5 degC; temperature dependent operation; Heart; Image reconstruction; Mice; Noise; Phantoms; Temperature dependence; Temperature measurement; Avalanche photodiodes; SPECT; biomedical nuclear imaging; scintillation detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2011.2165324
  • Filename
    6031785