• DocumentCode
    1337320
  • Title

    Planar double delay-line readout technique for microchannel plate detectors

  • Author

    Lampton, M. ; Siegmund, O.H.W. ; Raffanti, R.

  • Author_Institution
    Space Sci. Lab., California Univ., Berkeley, CA, USA
  • Volume
    37
  • Issue
    4
  • fYear
    1990
  • fDate
    8/1/1990 12:00:00 AM
  • Firstpage
    1548
  • Lastpage
    1549
  • Abstract
    A two-dimensional position-sensitive electrical readout system for microchannel plate detectors in spectroscopic applications has been devised. Two adjacent parallel coplanar delay lines are used. The difference in time of arrival of the electron event signal between the ends of the delay lines gives one coordinate of the centroid of each event pulse. The ratio of the event amplitudes on the two delay lines gives the orthogonal coordinate by means of charge partitioning. The performance characteristics of operational double delay lines for systems of this type are presented
  • Keywords
    delay lines; electron multipliers; phototubes; position sensitive particle detectors; charge partitioning; electron event signal; event amplitudes; microchannel plate detectors; parallel coplanar delay lines; planar double delay line readout; position-sensitive electrical readout system; two-dimensional; Anodes; Bandwidth; Delay lines; Delay systems; Detectors; Microchannel; Propagation delay; Spatial resolution; Spectroscopy; Timing;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.55868
  • Filename