• DocumentCode
    1337981
  • Title

    Dry calibration procedure of electromagnetic flowmeter for open channels

  • Author

    Michalski, Andrzej

  • Author_Institution
    Inst. of the Theory of Electr. Eng. & Electr. Meas., Warsaw Univ. of Technol., Poland
  • Volume
    49
  • Issue
    2
  • fYear
    2000
  • fDate
    4/1/2000 12:00:00 AM
  • Firstpage
    434
  • Lastpage
    438
  • Abstract
    This paper presents a simple approach to a dry calibration procedure for the electromagnetic flowmeter for both conducting and nonconducting open-flow channels. The author focuses on two main problems: variable water level in the measurement zone and variable water or ground conductivity and their influence on a flow signal. Based on the existing numerical model of electromagnetic flowmeter, the author describes the main idea of transfer function calculation with a few exemplary results. As an approximation instrument for a final form of correction function, the sigmoidal two-layer neural network has been applied
  • Keywords
    calibration; channel flow; electromagnetic devices; flowmeters; transfer functions; Poisson equation; backpropagation; conducting open-flow channels; correction factors; correction function; dry calibration procedure; electromagnetic flowmeter; nonconducting open-flow channels; numerical model; sigmoidal two-layer neural network; transfer function calculation; variable ground conductivity; variable water conductivity; variable water level; Calibration; Coils; Conductivity; Density measurement; Electrodes; Electromagnetic measurements; Fluid flow measurement; Magnetic field measurement; Magnetic liquids; Ultrasonic variables measurement;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.843092
  • Filename
    843092