• DocumentCode
    1344423
  • Title

    Design of a miniaturized ultrasonic bladder volume monitor and subsequent preliminary evaluation on 41 enuretic patients

  • Author

    Petrican, Paul ; Sawan, Mohamad A.

  • Author_Institution
    Dept. of Electr. & Comput. Eng., Ecole Polytech. de Montreal, Que., Canada
  • Volume
    6
  • Issue
    1
  • fYear
    1998
  • fDate
    3/1/1998 12:00:00 AM
  • Firstpage
    66
  • Lastpage
    74
  • Abstract
    Nocturnal incontinence (enuresis) affects 20% of children over four years old, and this figure typically decreases by 15% each year. At the age of 18, 1% of those people remain enuretic. Nocturnal enuresis can be treated by means of a conditioning device that awakens the patient once the urine level has reached a preestablished threshold of the capacity of his or her bladder. The authors have designed and implemented a portable miniaturized ultrasonic monitor, which permits estimation of the urine volume with an accuracy of 75%. Prototypes have been completed and validated on 41 patients (children) at Ste. Justine Hospital (Montreal). On the first group of 33 patients, the authors used a hand-held transducer to determine the volume detection range, which is accurate for volumes between 40 and 400 mL. With the second group of 8 patients, the device was mounted on an elastic belt around the abdomen. Measurements were taken in order to validate the accuracy of urine threshold detection and the activation of the corresponding alarm
  • Keywords
    alarm systems; biomedical equipment; biomedical ultrasonics; patient monitoring; ultrasonic equipment; volume measurement; 18 y; 4 y; Ste. Justine Hospital; alarm activation; bed wetting; bladder capacity; enuretic patients; hand-held transducer; medical instrumentation; miniaturized ultrasonic bladder volume monitor design; nocturnal enuresis; nocturnal incontinence; urine level; urine threshold detection accuracy; wakeup alarm; Ambient intelligence; Bladder; Hospitals; Medical treatment; Patient monitoring; Pediatrics; Prototypes; Ultrasonic imaging; Ultrasonic transducers; Ultrasonic variables measurement;
  • fLanguage
    English
  • Journal_Title
    Rehabilitation Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1063-6528
  • Type

    jour

  • DOI
    10.1109/86.662622
  • Filename
    662622