Design Guidelines for a Double Scattering Compton Camera for Prompt- \\gamma Imaging During Ion Beam Therapy: A Monte Carlo Simulation Study
Author :
Richard, M.-H. ; Chevallier, M. ; Dauvergne, D. ; Freud, N. ; Henriquet, P. ; Le Foulher, F. ; Létang, J.M. ; Montarou, G. ; Ray, C. ; Roellinghoff, F. ; Testa, E. ; Testa, M. ; Walenta, A.H.
Author_Institution :
Univ. de Lyon, Lyon, France
Volume :
58
Issue :
1
fYear :
2011
Firstpage :
87
Lastpage :
94
Abstract :
In hadrontherapy in order to fully take advantage of the assets of the ion irradiation, the position of the Bragg peak has to be monitored accurately. Here, we investigate a monitoring method relying on the detection in real time of the prompt γ emitted quasi instantaneously during the nuclear fragmentation processes. Our detection system combines a beam hodoscope and a double scattering Compton camera. The prompt-γ emission points are reconstructed by intersecting the ion trajectories given by the hodoscope and the Compton cones reconstructed with the camera. We propose here to study in terms of point spread function and efficiency the theoretical feasibility of the emission points reconstruction with our set-up in the case of a photon point source in air. First we analyze the nature of all the interactions which are likely to produce an energy deposit in the three detectors of the camera. It is underlined that upper energy thresholds in both scatter detectors are required in order to select mainly Compton events (one Compton interaction in each scatter detector and one interaction in the absorber detector). Then, we study the influence of various parameters such as the photon energy and the inter-detector distances on the Compton camera response. These studies are carried out by means of Geant4 simulations. We use a source with a spectrum corresponding to the prompt-γ spectrum emitted during the carbon ion irradiation of a water phantom. In the current configuration, the spatial resolution of the Compton camera is about 6 mm (Full Width at Half Maximum) and the detection efficiency 10-5. Finally, provided the detection efficiency is increased, the clinical applicability of our system is considered.
Keywords :
radiation therapy; solid scintillation detectors; Bragg peak; Compton events; Geant4 simulations; Monte Carlo simulation study; beam hodoscope; carbon ion irradiation; double scattering Compton camera; inter-detector distances; ion beam therapy; ion trajectories; monitoring method; nuclear fragmentation processes; photon energy; photon point source; prompt-gamma emission points; prompt-gamma imaging; prompt-gamma spectrum; spatial resolution; Compton camera; Geant4; hadrontherapy; ion beam therapy; prompt gamma;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2010.2076303
Filename :
5595525
Link To Document :
بازگشت