• DocumentCode
    1345052
  • Title

    Dynamic calibration of an aerosol monitor with natural and artificial alpha-emitters

  • Author

    Klett, A. ; Reuter, W. ; De Mey, L.

  • Author_Institution
    EG&G Berthold, Bad Wildbad, Germany
  • Volume
    44
  • Issue
    3
  • fYear
    1997
  • fDate
    6/1/1997 12:00:00 AM
  • Firstpage
    804
  • Lastpage
    805
  • Abstract
    Measurement of airborne artificial α-activity concentrations in the presence of naturally occurring radionuclides requires reliable subtraction of the natural contribution and precise calibration. Therefore, the commercially available aerosol monitor LB150D-R was calibrated dynamically at the IPSN/CEA in Saclay, France. Aerosols with activity median aerodynamic diameters of 0.4 μm and 4 μm labelled with 239Pu were generated and isokinetically sampled by the aerosol monitor. The response to 239Pu activity concentrations and detection limits in the presence of radon progeny were determined. The retention of aerosols and their distribution on the filter was investigated using fluoresceine labelled aerosols
  • Keywords
    aerosols; alpha-particle detection; calibration; proportional counters; radiation monitoring; radioactivity measurement; 0.4 mum; 4 mum; 239Pu; LB150D-R; Pu; activity median aerodynamic diameters; aerosol monitor; airborne artificial α-activity concentrations; dynamic calibration; filter; fluoresceine labelled aerosols; naturally occurring radionuclides; radon progeny; Aerodynamics; Aerosols; Calibration; Filters; Instruments; Manufacturing; Monitoring; Nuclear power generation; Radiation detectors; Sampling methods;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.603756
  • Filename
    603756